VHP gasified hydrogen peroxide sterilization watsa taga kuma aka sani da (VHP sterilization watsa taga) sterilization watsa taga, VHP irin zafi sterilization watsa taga, watsa majalisa, watsa akwati.
VHP gasified hydrogen peroxide sterilization watsa taga-VHP sterilization watsa taga-sterilization sterilization watsa taga samfurin gabatarwa da kuma amfani: VHP hydrogen peroxide watsa taga ne wani sterilization kayan aiki da zai iya samar da halitta decontamination a kan abu farfajiyar yayin watsa abubuwa tsakanin ayyuka daban-daban. VHP sterilization watsa taga Yana sanye da keɓaɓɓun hydrogen peroxide janareta, sterile iska samar da tsarin, electromagnetic ƙofar sarkar tsarin, rufe tsarin, sterilization bayan kawar da ragowar tsarin, musamman HMI (mutum-inji dubawa) da sterilization kafofin watsa labarai samar da tsarin. Ana amfani da taga mai watsawa na hydrogen peroxide sosai don sterilization na yanayin zafin jiki na farfajiyar magunguna, likita, tsaftacewa, gwajin halitta da sauransu.
VHP gasified hydrogen peroxide sterilization watsa taga-VHP sterilization watsa taga-sterilization sterilization watsa taga aiki ka'idar:
VHP sterilization watsa taga amfani da hydrogen peroxide plasma a yanayin zafin jiki gas fiye da ruwa da kuma tururi yanayin mafi da siffofin kashe spores ikon, bayan samar da kyauta H2O2﹢da H2O2 -, da aka yi amfani da su kai hari cell abubuwa, ciki har da lipids, furotin da DNA tissues, lalata maɓallin haɗin, cimma cikakken sterilization manufa, VHP sterilization watsa taga amfani da musamman sterilization kafofin watsa labarai bayar da tsarin sa hydrogen peroxide plasma mafi daidai, cimma sterilization sakamakon.
Dehumidification → Sterilization → cire ragowar → Bincika aiki yanayin
Dehumidification - sa iska zagaye a cikin kewaye ta hanyar dehumidification na'urar, rage dangi zafi a cikin kewaye, sa VHP sterilization inganci high;
Sterilization - Shigar da tururi na hydrogen peroxide da kuma kiyaye matattarar da ake tsammani a cikin keɓaɓɓen, kiyaye matattarar VHP sama da 700PPM da kuma kiyaye lokacin sterilization sama da 30min;
Raba ragowar - dakatar da shigar da hydrogen peroxide gas, canza sterilization tsarin zuwa kawar da ragowar tsarin, sa hydrogen peroxide gas zagaye a cikin keɓaɓɓen ya rushe ta hanyar catalyst, rage hydrogen peroxide matakan a kasa da 10ppm, sa'an nan kuma ta hanyar iska don cimma ragowar darajar ba fiye da 1ppm.
Bincika aiki jihar - tsarin canzawa zuwa tsabta kulawa jihar, bisa ga saita aiki iska gudun da kuma m matsin lamba a cikin gidan, atomatik misali canzawa daidaita samar da iska yawa, return iska yawa, sabon iska yawa, kiyaye a cikin gidan tsabta da kuma m matsin lamba jihar, yayin da online gano aiki yanki tsabta, kuma za a iya hannu fara planktonic samfurin.
VHP gasified hydrogen peroxide sterilization watsa taga-VHP sterilization watsa taga-sterilization sterilization watsa taga fasaha halaye:
1, atomatik gano leakage: atomatik inflation a kan cabinet, gano ko tsarin ya kasance rufe, atomatik gano sealability.
2, sauri steaming ruwa hydrogen peroxide zuwa sub-micron matakin particles, sa'an nan kuma ta hanyar plasma janareta tsari a matsayin gas na plasma hydrogen peroxide, inganta kashe ƙwayoyin cuta inganci.
3, nanodegradation tace iya canza plasma halin hydrogen peroxide zuwa ruwa tururi da oxygen, ban da ragowar lokaci kawai dauki 10 ~ 20min.
4, cikakken atomatik mai hankali iko, a kan iko tsari rikodin, kowane sassa aiki yanayin nuni, rahoto fitarwa, ƙararrawa bayani.
5, abu rack ne irin drawer, ga kowane irin marufi abubuwa suna da keɓaɓɓun rack, tabbatar da kowane irin samfurin waje farfajiyar da yawa fallasa a cikin iska, taimakawa sterilization.
6, ƙofar interlocking aiki, bayan ragowar karshen kawai za a iya buɗe tsabtace matakin ƙofar, amfani da inflatable rufe ƙofar, tabbatar da hatimi.
7, amfani da low ikon amfani da shigo da EC iska, makamashi ceton inganci.
8, VHP mayar da hankali, iska gudun, matsin lamba bambanci, zafi da zafi, da kuma samar da curves da rahotanni. Cikakken takardar shaidar kwamfuta.
VHP gasified hydrogen peroxide sterilization watsa taga-VHP sterilization watsa taga-sterilization sterilization watsa taga fasaha halaye:
1, kayan: 316 bakin karfe zane-zane.
2, rufewa: cabinet ne a karkashin matsin lamba na 500Pa, a cikin 10min leakage rate ba fiye da 5%.
3, iska gudun: 0.36 ~ 0.54m / s, iska gudun daidaitawa, kula da 20 ~ 60Pa lokacin sterile aiki, matsin lamba a karkashin tsayayye yanayi kula da saiti darajar ba kasa da 10Pa.
4, atomatik cire ragowar inganci: mayar da hankali <1PPM.
5, tsabtace: Class A.
6, wutar lantarki: AC220V ± 22V, 50Hz ± 1Hz.
7, tabbatar da abun ciki: sterilization inganci: halitta nuna alama, sinadarai gwaji takarda A-matakin stratification: iska gudun, matsa lamba bambanci, ƙura ƙwayoyi, plankton.
8, Size: Za a iya tsara bisa ga bukatun.
VHP gasified hydrogen peroxide sterilization watsa taga-VHP sterilization watsa taga-sterilization sterilization watsa taga yi sigogi:
Wutar lantarki: AC 220V / 50Hz
ikon: 2000W
Ƙara gudun: 1 ~ 3g / min
Gas daukar tsarin: matsa iska: 0.3-0.6Mpa; Gas kwarara: < 6 m3 / h
Gasification zafin jiki: ≤100 ℃
hayaniya: ≤70dB (A)
Sterilizer: 35% abinci grade hydrogen peroxide mafita
Kashewa Rate: Kashewa ikon a kan thermophilic fat germs kai 6 log
Yanayin aiki: Manual ko atomatik
Sterilization zagaye: 90 minti (kimanin)
VHP matattarar lokacin sterilization: > 700ppm;
VHP ragowar matakin kasa da daidai da 1ppm.
Cibiyar zafin jiki kewayon ne 30 ℃ ~ 35 ℃, zafi kewayon ne 40% RH ~ 50% RH.
Real-lokaci sa ido a cikin dakin zafin jiki, zafi, matsin lamba bambanci, VHP mayar da hankali da kuma sanye da data ajiya buga aiki, ta amfani da launi-band takarda firintar. Bude ƙofar ƙararrawa, kulle ƙofar ƙararrawa aiki, biyu ƙofar kulle aiki.
VHP gasified hydrogen peroxide sterilization watsa taga-VHP sterilization watsa taga-sterilization sterilization watsa taga bayanan sigogi:
VHP gasified hydrogen peroxide sterilization watsa taga-VHP sterilization watsa taga-sterilization sterilization watsa taga aikin tafiya:
Leakage gwajin kafin fara sterilization zagaye - rufe duk bawul a cikin sterilization dakin, da matsin lamba tashi yanayi ta hanyar inji matsin lamba mita lura, idan matsin lamba tashi kudin ya dace da tsarin ka'idodin, to, tsarin da aka rufe, za a iya fara pre-processing
1. Da farko bude inji famfo a kan sterilization kamfanin pumping inji, cire ruwa da danshi.
2. Bude iska bawul, rush a cikin bushewa, sterile iska.
3. Lokacin da matsin lamba ya kai yanayi matsin lamba, bude inji famfo sake, kara bushewa
5. allura hydrogen peroxide a cikin dakin, saboda high inji a cikin pre-processing lokaci dakin, sa hydrogen peroxide samun cikakken yaduwa, sosai inganta sterilization sakamakon.
6. Bayan karshen sterilization, injin famfo bude, sa a cikin gidan iska matsin lamba sauka zuwa 1Torr
7. Bude iska bawul, rush a cikin bushe, sterile iska, lokacin da matsin lamba a cikin gida kusa da al'ada matsin lamba, inji famfo fara aiki, fitar da iska a cikin gida, da wadannan ragowar gas a karkashin wani catalyst aiki a cikin m oxygen da ruwa.
8. iska lokaci karshen, bude iska bawul, sabon zagaye na sterile iska sake allura a cikin sterilization tank don matsin lamba dawo da al'ada, a sterilization zagaye karshen.